High Voltage Digital X Ray Machines X-ray Glasses

  Product Description32kw 400mA Floor-Mounted Digital Radiography X-Ray machines No.ComponentPerformance parameter1Medical diagnostic X-ray high-frequency high-voltage generatorOutput Power: 32kWInput Voltage: 220VAC,±10%,50HzInverter Frequency: 25kHz,±10%kV range: 40 kV~150kVmA range: 10 mA~400 mAms range: 1ms~6300 msmAs range: 0.1mAs~150mAsTech mode: kV-mA(AEC);kV-mAs;kV-mA-msAPR: 2000AEC(Optional):  AEC Ionization chamber power supply:+/-15VMinimum exposure time:5msAuxiliary power supply: +24V DC 2A;24V AC 8ACommunication Interface: RS-2322X-ray Tube AssemblyX-ray Tube Voltage Range: 40kV~125kVNominal Focal Spot Value: 0.6mm(Small Focus)/1.2mm(Large Focus)Nominal Anode Input Power: 20kW(Small Focus)/40kW(Large Focus)Anode Heat Content: 150KHu3Flat Panel DetectorImage Sensor: a-Si (Amorphous Silicon) TFTScintillator: CsI:TlEffective Array: 3072 ×3072Pixel Size: 139 umEffective Area (H x V): 427mm×427mmPreview Time: 3sImaging Time: 5sSpatial Resolution: 3.6 lp/mmAD Conversion: 16 bitData Interface: Gigabit EthernetTrigger Mode: AED(optional)/Software/PrepDimension (L × W × H): 460mm×460mm×15mmWeight: 4kgPower Consumption: 20W4CollimatorControl Mode: Manual controlMaximum Radiation: 150kVInherent Filtration Al. Equivalent: 1mm ALRectangular X-ray Field Selection Set Focus Distance, SID=100cm: Min 2cm×2cm;Max 43cm×43cmX-Ray Field Deviation: ≤1% of SIDLight Field Indicator @ SID=100cm: ≥180luxLight Field Indicator Precision: ≤1% of SIDLight source ON time: 30sMaximum Radiation Leakage EN 60601-1-3: